在线离线TOC分析仪 纯水总有机碳分析仪 测定原理:仪器采用紫外氧化的原理,将水溶液中的总有机碳氧化为二氧化碳,并且测定其含量,二氧化碳的测试采用的是直接电导率法,通过测试经过氧化反应的水溶液总碳含量和未经过氧化反应的样品总无机碳的含量差值来测定总有机碳含量,即:总有机碳(TOC)=总碳(TC)-总无机碳(TIC)。
在线离线TOC分析仪 纯水总有机碳分析仪 北广公司仪器研发部拥有行业内 优秀的、权威的研发力量及技术团队,长期聘请 精密仪器系的专家为技术顾问,并成立新型材料检测仪器研发中心,是推动公司材料检测技术的提高和试验方法创新的重要技术保证力量。
Determination principle: the instrument adopts the principle of ultraviolet oxidation to oxidize the total organic carbon in the aqueous solution into carbon dioxide and determine its content. The carbon dioxide test adopts the direct conductivity method. The total organic carbon content is determined by testing the difference between the total carbon content of the aqueous solution after oxidation reaction and the total inorganic carbon content of the sample without oxidation reaction, that is, total organic carbon (TOC) = total carbon (TC) - Total inorganic carbon (TIC).
The instrument R & D Department of BG has the most outstanding and authoritative R & D force and technical team in the industry. It has long hired experts from the precision instrument department of Tsinghua University as technical consultants and established a new material testing instrument R & D center, which is an important technical guarantee to promote the improvement of the company's material testing technology and the innovation of test methods.
主要特征:
1、高精度、高灵敏度,操作简单。
2、人性化操作界面,有一键运行功能,自动管路清洗功能。
3、高性能CPU,触摸屏设计,超大640*480点阵真彩显示器。
4、不用拆开机箱更换UV灯和泵管。
5、检测上限可设定,自动上限报警功能。
6、具有RS232数据接口,历史数据可存储6个月。
7、离线检测和在线检测可选配。
8、具有打印功能
Main features:
1. High precision, high sensitivity and simple operation.
2. Humanized operation interface, with one key operation function and automatic pipeline cleaning function.
3. High performance CPU, touch screen design, large 640*480 dot matrix true color display.
4. Do not disassemble the chassis to replace the UV lamp and pump pipe.
5. The detection upper limit can be set, and the automatic upper limit alarm function.
6. With RS232 data interface, historical data can be stored for 6 months.
7. Offline detection and online detection are optional.
8. With printing function
enclosure:
Test host One
Acquisition tube Two
power cord One
Control system a set
data acquisition system a set
Product operation manual One copy
Product certificate and warranty card One copy
附件:
试验主机 一台
采集管 两条
电源线 一条
控制系统 一套
数据采集系统 一套
产品使用说明书 一份
产品合格证、保修卡 一份
离线检测
离线检测时,仪器可从样品瓶或其它没有压力的容器直接取样。仪器管路冲洗和仪器校准应在离线状态操作。若样品中有不可溶性微粒,应经过滤膜(孔径60μm或更小)过滤后进入仪器,以防止样品中的微粒阻塞仪器。仪器的进样管为1/16英寸的Teflon管,经蠕动泵抽进管路中的水样流速约为0.5 ml/min。
Offline detection
During off-line detection, the instrument can directly sample from the sample bottle or other containers without pressure. Instrument pipeline flushing and instrument calibration shall be operated offline. If there are insoluble particles in the sample, it shall pass through the filter membrane (aperture 60 μ M or less) enter the instrument after filtration to prevent particles in the sample from blocking the instrument. The sampling tube of the instrument is a 1/16 inch Teflon tube, and the flow rate of the water sample pumped into the pipeline by the peristaltic pump is about 0.5 ml / min.
Carbon dioxide measurement cycle
The instrument detects one data every 4 minutes (including TOC value and conductivity value). TC and tic are detected independently in the 4-minute measurement cycle.
二氧化碳测量循环
仪器每4分钟检测得出一个数据(包括TOC值和电导率值),在4分钟的测量循环中TC和TIC是独立检测的。
使用与操作方法
开始检测并计入检测次数,检测完毕后显示最后一次检测结果,其中几次检测的结果均自动保存在查询记录当中。检测次数的设置方式为:用“选择”键移动光标,用“设置”键修改数字,按“确定”键进行确认,进入分析界面。先进行四次冲洗过程
Usage and operation method
Start detection and count the detection times. After detection, the last detection result will be displayed. The results of several tests will be automatically saved in the query record. The setting method of detection times is as follows: use the "select" key to move the cursor, use the "set" key to modify the number, press the "OK" key to confirm, and enter the analysis interface. Conduct four flushing processes first